MyWorldGo Important notes for Glovebox Circulation

Blog Information

  • منشور من طرف : qqyouc qqyouc
  • نشر على : Jun 20, 2025
  • الآراء : 42
  • الفئة : جنرال لواء
  • وصف : Set the gas cylinder pressure to 60-80 psi. Insufficient pressure may prevent the initiation or proper maintenance of circulation.

نظرة عامة

  • Set the gas cylinder pressure to 60-80 psi. Insufficient pressure may prevent the initiation or proper maintenance of circulation.

    When the antechamber is not in use, maintain a slight negative pressure and ensure both "Evacuation" and "Refill" are in the "off" position.

    Circulation cannot be initiated during the Purging or Regeneration processes.

    The circulation blower automatically adjusts its operating frequency based on moisture and oxygen levels, typically running at either 15 Hz or 30 Hz.

    Glovebox Circulation

    If moisture and oxygen levels do not decrease after an extended period of circulation, the purifier may require regeneration, or you may need to contact a VTI service engineer for support.